GRUBATEC AG Measurement & Control Technology

Model MH 3156 Handheld Pressure Gauge with Interchangeable Probes

Universal two-channel pressure gauge for interchangeable sensors up to 1,000 bar

MH 3156 Handheld Pressure Gauge with Two Interchangeable Pressure Sensors

The SIKA MH 3156 is a versatile, two-channel digital pressure gauge designed for both mobile and stationary measurement applications. Unlike a pressure gauge with a fixed, built-in measurement range, the MH 3156 uses external, interchangeable pressure sensors. This allows the same handheld gauge to be used for a wide variety of pressure ranges and types of pressure.

A distinctive feature of the MH 3156 is its two sensor connections. Two compatible pressure sensors can be connected and read simultaneously. In addition to the two individual readings, the device can directly calculate and display the pressure difference between Sensor 1 and Sensor 2.

Depending on the sensor used, measurement ranges are available from very low mbar levels up to 1,000 bar. This makes the system suitable for both precise low-pressure measurements and industrial high-pressure applications.

A handheld pressure gauge for a wide range of pressure measurement ranges

With the MH 3156, the actual pressure measurement range is determined by the connected pressure sensor. This allows the user to combine different sensors with the same data logger.

The available sensors cover, among other things, relative pressure, absolute pressure, negative pressure, differential pressure, atmospheric pressure (or barometric measurements), and vacuum measurements. The type of measurement available depends on the specific sensor connected.

This modular design is particularly useful when different pressure ranges need to be measured regularly at a workstation and a separate measuring device is not to be used for each application.

Two sensor connections for simultaneous measurements

The MH 3156 has two 6-pin sensor connectors. Both sensors can be connected at the same time.

The main display shows the reading from Sensor 1. The secondary display can show either the reading from Sensor 2 or the calculated difference between Sensor 1 and Sensor 2.

This makes the MH 3156 particularly suitable for applications in which two pressure points need to be monitored simultaneously or compared directly with one another.

Direct Differential Pressure Measurement

One of the MH 3156's most important features is differential pressure measurement using two sensors. The device automatically calculates the difference between Sensor 1 and Sensor 2 and displays it directly.

This makes it possible, for example, to analyze pressure losses, pressure differences between two measurement points, or differential pressures in systems without having to calculate the values separately afterward.

Automatic Sensor Detection

The connected pressure sensors are automatically detected by the handheld meter. The corresponding sensor data is stored in the sensor and is read by the MH 3156.

This ensures that the appropriate measurement range is automatically taken into account when the sensor is replaced. This enables convenient plug-and-play operation with compatible SIKA pressure sensors.

Measurement ranges from the mbar range up to 1,000 bar

A wide selection of different pressure sensors is available for the MH 3156. Depending on the sensor, the measurement ranges extend from a few millibars up to 1,000 bar.

Sensors with very high resolution are available for measuring particularly small pressure differentials. For industrial high-pressure measurements, stainless steel pressure sensors with measurement ranges up to 1,000 bar are available.

Measuring Different Types of Pressure

When equipped with the appropriate sensors, the MH 3156 can measure various types of pressure. These include, among others, positive pressure, negative pressure, absolute pressure, and differential pressure.

With absolute pressure sensors, the device can also be used for air pressure and barometric measurements. A built-in sea-level correction feature allows the measured air pressure to be converted to sea level.

Switchable Print Units

The MH 3156 supports various pressure units. Depending on the connected sensor, you can select units such as mbar, bar, Pa, kPa, MPa, mmHg, PSI, and meters of water column, among others.

This allows the measuring device to be flexibly adapted to different applications, industries, and international measurement standards.

Slow and Fast Pressure Measurements

The MH 3156 offers various measurement modes for different applications.

In the standard “SLOW” mode, the device operates at a measurement frequency of 4 Hz. This mode is particularly suitable for static pressures, slow pressure changes, air pressure measurements, and leak tests.

For dynamic processes, the “FAST” and “Peak Detection” measurement modes are available. In these modes, the device operates at a measurement rate of more than 1,000 measurements per second.

Peak Detection for Rapid Pressure Spikes

In peak detection mode, the measurement signal is processed unfiltered at a frequency of over 1,000 Hz. This allows for the detection of very short pressure spikes and rapid pressure changes.

This feature is particularly well-suited for applications in which the maximum or minimum pressure of a fast-moving process needs to be measured.

FAST Mode for Dynamic Measurements

Even in FAST mode, the MH 3156 performs more than 1,000 measurements per second. Unlike peak detection, however, the signal is filtered.

As a result, the device responds quickly to pressure changes and is at the same time less sensitive to very short transient spikes.

Logger Function for Recording Measurement Data

The MH 3156 features a built-in data logger. Measurement values can be recorded either manually or automatically at preset intervals.

In manual storage mode, up to 99 data sets can be stored. A cyclic logger is available for automatic recording.

The measurement interval can be adjusted flexibly, allowing for both short-term measurement series and longer-term pressure monitoring.

Min, Max, and Hold Functions

During a measurement, the minimum and maximum recorded values can be saved and displayed at any time.

The Hold function allows you to freeze the current measurement value on the display. This makes it easy to document or read individual measurement points.

Tare and Zero Function

The tare function allows you to set the current reading as a new reference point. All subsequent readings are then displayed relative to this value.

In addition, a permanent zero-point calibration is possible to correct minor zero-point deviations in the sensor.

Averaging for Stable Measurement Readings

For fluctuating pressure signals, averaging can be enabled. The MH 3156 calculates an averaged measurement value over a configurable time period.

This feature is particularly helpful when a process exhibits slight pressure fluctuations and a stable, representative measurement is required.

Alarm functions for both sensors and differential pressure

Individual upper and lower alarm limits can be configured for Sensor 1 and Sensor 2. In addition, custom limit values can be defined for the calculated pressure differential.

Depending on the configuration, exceeding or falling below a limit can be indicated visually and audibly.

PC Connection and Analog Output

The device output can be used as a serial interface or as an analog output.

Using a suitable interface adapter, the device can be connected to an RS232 or USB port. This allows measurement data to be transferred to a computer for further analysis.

Alternatively, a freely scalable analog output ranging from 0 to 1 V is available. You can specify whether to output Sensor 1, Sensor 2, or the difference between the two sensors.

Compact and designed for mobile use

The housing is made of impact-resistant ABS plastic and features a membrane keyboard and a transparent protective screen. The front panel is IP65-rated.

With dimensions of just 142 × 71 × 26 mm and a weight of approximately 150 g, the MH 3156 is easy to use on the go.

Power is supplied by a 9-V block battery. Alternatively, an optional AC adapter can be used.

Optional Calibration Certificates

For applications involving documentation and quality control, the MH 3156 can be supplied with an optional DAkkS calibration certificate or a factory test report.

Typical Applications of the MH 3156

  • Differential Pressure Measurements
  • Pressure Comparison Between Two Measurement Points
  • Leak Tests
  • Measurement of Static and Dynamic Pressures
  • Detection of Rapid Pressure Spikes
  • Monitoring Pressure Losses
  • Service and Maintenance
  • Machinery and Plant Engineering
  • Laboratory and Testing Applications
  • Air Pressure and Barometer Measurements Using an Appropriate Sensor

The Key Benefits of the MH 3156

  • Two sensor connections
  • Direct Differential Pressure Measurement
  • Replaceable Pressure Sensors
  • Automatic Sensor Detection
  • Measurement ranges up to 1,000 bar, depending on the sensor
  • Various printing methods using a single base unit
  • Measurement frequency above 1,000 Hz for dynamic processes
  • Peak Detection for Rapid Pressure Peaks
  • Built-in logging function
  • Min/Max, Hold, Tare, and Average functions
  • Alarm functions for both sensors and differential pressure
  • PC interface or freely scalable analog output
  • Compact and rugged design

Questions and Answers About the MH 3156 Handheld Pressure Gauge

What is the MH 3156?

The MH 3156 is a digital, two-channel handheld meter for interchangeable SIKA pressure sensors. It can read two sensors simultaneously and also calculate the pressure difference between them.

What is the main difference between this and a standard digital pressure gauge?

The pressure sensor is not permanently integrated into the MH 3156. Different sensors can be connected, allowing the same data logger to be used for various pressure ranges and pressure types.

How many pressure sensors can be connected?

The MH 3156 has two sensor connections and can read data from two compatible pressure sensors simultaneously.

Can the MH 3156 measure differential pressure?

Yes. The device directly calculates the difference between Sensor 1 and Sensor 2 and can display this value on the secondary display.

What is the maximum pressure that can be measured?

The maximum measurement range depends on the connected sensor. The available range of sensors includes measurement ranges up to 1,000 bar.

Can very low pressure also be measured?

Yes. Sensors in the mbar range are available for low-pressure applications. The smallest ranges listed in the data sheet start at just a few millibars.

What types of pressure can be measured?

Depending on the sensor, measurements such as relative pressure, negative pressure, absolute pressure, differential pressure, atmospheric pressure, and vacuum are possible.

Does the device automatically detect the connected sensor?

Yes. Compatible SIKA sensors are automatically detected, and the corresponding measurement range setting is applied.

Is it possible to use two different measurement ranges at the same time?

The MH 3156 can process data from two compatible sensors simultaneously. However, for direct differential measurement, appropriate combinations of sensors and measurement ranges should be selected.

How fast does the MH 3156 measure?

In standard mode, the measurement frequency is 4 measurements per second. In FAST and Peak Detection modes, the device operates at more than 1,000 measurements per second.

What is peak detection used for?

Peak detection is used to detect rapid pressure spikes and dynamic pressure changes. The unfiltered signal is sampled at over 1,000 Hz.

Does the MH 3156 have a data logger?

Yes. Measurement values can be saved manually or automatically using the built-in logging function.

How many readings can be saved manually?

In manual logger mode, 99 data records can be stored.

Can the measurement interval be adjusted?

Yes. You can configure an automatic recording interval for the cyclic logger.

Can the MH 3156 sound an alarm?

Yes. Thresholds can be set for both sensors as well as for the pressure differential. Depending on the configuration, a visual and audible alarm is triggered.

Does the device have an analog output?

Yes. The device output can be operated as a freely scalable 0…1 V analog output.

Can the measured values be transferred to a PC?

Yes. Using a suitable galvanically isolated interface adapter, the device can be connected to an RS-232 or USB interface.

What printing units are available?

Depending on the sensor, units such as mbar, bar, Pa, kPa, MPa, mmHg, PSI, and meters of water column are available, among others.

Is the device suitable for mobile measurements?

Yes. The MH 3156 weighs approximately 150 g and is powered by a 9-V block battery as standard.

Which sensors are permitted for use?

According to the operating instructions, suitable sensors from the MSD or MXD series are recommended. The measuring device should be turned off when replacing the sensor.

Are calibration certificates available?

Yes. A DAkkS calibration certificate and a factory test certificate are available as options.

Technical data

Technical data

Type MH 3156
Device Type Dual-pressure handheld meter with interchangeable sensors
Sensor Connections 2 × 6-pin Mini-TC connectors with locking mechanism and automatic sensor detection
Measuring range Depending on the sensor, up to 1,000 bar
Types of Printing Overpressure, underpressure, differential pressure, absolute pressure, atmospheric pressure / barometer, and vacuum—depending on the sensor
Accuracy of the Base Unit ±0.1% of the final value, as specified in the data sheet
Units mbar, bar, Pa, kPa, MPa, mmHg, PSI, mH₂O
Default measurement rate 4 measurements per second
FAST / Peak Detection >1,000 measurements/s
Logger – Manual 99 records
Logger – cyclic 4,000 records according to the data sheet
Logging Interval 1 s…60 min
Additional Measurement Functions Peak value storage, averaging, sea-level correction, differential pressure measurement
Analog output 0…1 V, freely scalable
PC Interface Via galvanically isolated interface converters to RS-232 or USB
Display 2 × 4½-digit LCD
Display range -19,999…19,999 digits
Supply 9-V block battery or optional AC adapter
Housing Impact-resistant ABS plastic
Protection class IP65 on the front
Dimensions 142 × 71 × 26 mm
Weight about 150 g
Item Number EME8GMH3156000

Wide selection of pressure sensors

A wide range of interchangeable pressure sensors is available for the MH 3156. The measurement ranges extend from the high-resolution mbar range up to industrial high-pressure measurements of 1,000 bar. Depending on the sensor, relative or absolute pressure measurements, as well as measurements of various media, are possible.

We would be happy to assist you in selecting the right sensor for your measurement range, pressure type, medium, and desired accuracy.

Request

Request

This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
GRUBATEC Mess- und Regeltechnik