Test block Test equipment for checking the calibration accuracy for humimeter measuring devices with conductance measuring method

Test block for humimeter FL1, FL2, FLS, FLH, BLL, BL2, WLW, BLW, SLW, RM1 and BLO - Dimensions: 55 x 55 x 45 mm, weight: 96 g

Test block for humimeter FL1, FL2, FLH, BLL, BL2, WLW, BLW, SLW, RM1 and BLO - Dimensions: 55 x 55 x 45 mm, weight: 96 g
  • Quick check of the entire measuring range
  • detailed instruction
  • Simple handling

Description:

With the test block, you can easily and precisely check the calibration of your measuring device's lance. Using the "test block" characteristic makes the checking process quick and easy: you simply place the measuring tip on the test block. This test block allows you to test the calibration of the sensor at two specific points. If the displayed value deviates from the target range, we or our competent sales partners will be happy to help you further.

Our test blocks are made of high-quality materials to ensure a long service life and high reliability. By regularly checking and calibrating with the test block, you ensure that your measuring device is always accurate and reliable. This is particularly important in applications where precise measurements are crucial, such as quality control, research and development or industrial production.

Cleaning

Electrodes
- If the measuring head is dirty, it can be cleaned with alcohol.
Test block
- If the test block is dirty, it can be cleaned with a damp cloth.
cleaned with a damp cloth.

 

Care instructions

- Do not leave the test block in the rain, it is not waterproof.
- Do not expose the test block to extreme temperatures.
- Avoid strong mechanical shocks or loads.
- Store the test block in a climate of 30 to 70 % humidity

 

Positioning the test block

The two electrodes of the test block must rest firmly on the metal parts in front of and behind the black plastic insulator.

 

The sensor tube or the measuring tip must be held in the air during the check

Humimeter BLL

" The two electrodes of the test block must rest firmly on the metal parts in front of and behind the transparent plastic insulator.
" The sensor tube or the measuring tip must be held in the air during the test.

humimeter WLW, BLW & SLW

" Remove the nails and nuts.
" The two electrodes of the test block must each be held in the holes of the grub screws.
" The electrodes must be held with light pressure on the grub screws.
" The measuring head must be held in the air during the test.