Type TP 17ZERO TP Solid Dry Block Temperature Calibrator Dry block -10...100 °C Accuracy ±0,05 °C at 0°C

Compact design for on-site use / Optimized for ice point generation

Operating instructions for SIKA Type TP 17ZERO - Solid dry-block temperature calibrator

The SIKA TP 17ZERO dry-block temperature calibrator has been specially developed for temperature calibrations in the range from -10 °C to 100 °C. With its compact design, it is ideal for on-site use, but also for use in laboratories, workshops and measuring rooms. Its particularly high accuracy of ±0.05 °C at 0 °C makes it an ideal choice for creating ice reference points and for precise calibrations in the lower temperature range.

Features:

  • Compact design: Ideal for mobile use on site, easy to transport and flexible to use.
  • Optimized for the generation of ice reference points: Perfect solution for applications where precise zero point temperatures are required.
  • High accuracy: With an accuracy of ±0.05 °C at 0 °C, the TP 17ZERO guarantees precise measurement results.
  • Dry block function: This function enables quick and uncomplicated temperature calibration, regardless of viscosities, flash points or outgassing.

Areas of application:

The TP 17ZERO dry-block temperature calibrator is used in many industries and applications:

  • Installation and commissioning: For the precise calibration of temperature sensors during the installation of systems.
  • Manufacturing and production: Monitoring and calibration of sensors in production to ensure quality.
  • Maintenance and service: Regular calibration of measuring devices to ensure functionality.
  • Quality assurance and test equipment monitoring: Calibration to comply with measurement standards and ensure measurement accuracy in production.
  • Laboratory applications: Use for generating reference points and for fine adjustment of temperature measuring devices in the laboratory.

Mode of operation:

The dry block function of the TP 17ZERO has been specially developed to enable simple and precise calibration both in the laboratory and in the field. The calibrator ensures optimal thermal coupling between the block and the test object by using the correct transition sleeve. This sleeve should have an inner diameter of 0.5 mm larger than the outer diameter of the test object to ensure the best thermal connection.

Almost all straight temperature sensors with different lengths and diameters can be calibrated with the adapter sleeves. The dry block function covers the entire temperature range from -10 °C to 100 °C without the need to change the calibration medium. This enables efficient and time-saving calibration without having to take into account physical properties of the medium such as viscosities or flash points.

Advantages:

  • Quick and easy calibration: The dry block function enables precise and quick calibration, regardless of the physical properties of the medium.
  • High precision: The accuracy of ±0.05 °C at 0 °C ensures exact measurement results, especially when creating ice reference points.
  • Wide range of applications: Ideal for on-site use, laboratory applications as well as field calibrations in a variety of industries.
  • No need for special calibration media: The dry block function works regardless of viscosities, flash points or outgassing, which significantly extends the range of applications.

Technical data:

  • Temperature range: -10 °C to 100 °C
  • Accuracy: ±0.05 °C at 0 °C
  • Power supply:
    • 100...240 VAC, 50 / 60 Hz

Summary:

The SIKA TP 17ZERO dry-block temperature calibrator offers a precise, compact and reliable solution for the calibration of temperature sensors in the low temperature range. Especially the optimization for the generation of ice reference points and the high measurement accuracy make the TP 17ZERO the ideal choice for applications where highest precision and reliability are required. The ease of use and versatility of the calibrator help to increase efficiency and quality in a wide range of industrial applications.

 

 

Power supply
100...240 VAC, 50 / 60 Hz

 

Technical data

 

 

 

 

Request

This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.

Request