Type induQ VMM Magnetic Inductive flow sensors with flange connection Metal version Measuring range 0...18833.3 l/min DN 15...200

Magnetic-inductive flow sensor for large nominal diameters DN15...200 / No mechanical wear / Robust industrial design / Simple menu-driven operation / Programming via display / Supplied with factory test certificate

The electromagnetic flow sensors of the VMM series, called induQ®, are precise measuring devices for electrically conductive liquids and, due to their robust design, are also suitable for harsher ambient conditions.

The magnetic-inductive flow sensor type VMM from SIKA is used in the following areas, among others:

- Water and wastewater treatment

- Mining, cement and mineral industry

- Pulp and paper production

- Steel industry

- Energy industry and utilities

- Agribusiness

The intelligent flow sensors of the induQ® series operate according to the induction principle: The measuring tube is located in a magnetic field (B). When an electrically conductive medium flows through the measuring tube at right angles to the magnetic field, a voltage (U) is induced which is proportional to the average flow velocity and is tapped by two electrodes. Two variants are available for an output signal proportional to the flow:

- Frequency output signal

- Analog and frequency output signal

The pulse rate can be configured at the factory depending on the type.

<td „>VMM40

TYPE VMM15  VMM25 VMM32 VMM 50 VMM65 VMM 80 VMM100 VMM125 VMM150 VMM200
Characteristics
Nominal width DN 15 DN25 DN32 DN40 DN 50 DN 65 DN 80 DN 100 DN 125 DN 150 DN 200
Process connection Flange connection according to EN 1092-1,JS B2220 IOK or ANSI B16.5
Inner diameter hard gurnmi 14,0 27,0 33,3 38,0 48,5 64,3 76,9 102,5 127,7 156,3 205,1
PTFE 14,0 27,0 33,3 38,0 48,5 63,3 75,9 102,5 124,7 152,3 201,1
Measuring range flow velocity [m/s] 0...10
Volume flow [m3/h] 0...6,3 0...17,6 0...28,9 0...28,9 0...45,2 0...70,6 O ...119,4 0...180,9 0...282,7 0...441,7 0...636,1 0...1130
Messgenauigkeit* v = 1…10 m/s v< 1 m/s zusätzlich Frequenzausgang Analogausgang ±0.5 % of measured value ±0.4 % of measured value ±0.05 % per 10 K ±0.1 % per 10 K
Repeatability ±0,15 %
Response time < 100 ms (Abhängig von den Einstellungen der Elektronik)
Signal output from O m/s
Measuring medium Water and other conductive liquids
Min. conductivity of the measured medium 20 pS/cm
Media temperature
Hard rubber 0...90 °C
PTFE process connections process connections -20...100 °C at 40 bar -20...130 °C at 25 bar -20...180 C at 16 bar (only for separated version) Min. -10 cc (steel) Min. -20 o c (stainless steel)
Ambient temperature hard rubber 0...80 °C
PTFE -20...100 °C
process connections process connections display electronics Min. -10 °C (steel) Min. -20 o c (stainless steel) -20...50 C (Below 0 °C the readability of the LCD display is limited)
Transport / storage temperature -20...60 °C
Compressive strength
.. EN 1092-1 PN 40 PN 40 PN 40 PN 40 PN 40 P N 16 PN 40 PN 16 PN 40 PN 16 PN 40 PN 16 PN40 PN 16 PN 40 PN 10 PN 16 PN 25 PN 40
JS B2220 IOK ANSI B16.5 150 RF 9.8 bar
19.6 bar (process connection steel)
15.9 bar (stainless steel process connection)
Display LCD two lines, backlight
Operation 6 keys, menu-driven
Protection class EN 60529 IP67

 

Measuring range - volume flow
0...105 l/min
0...293.3 l/min
0...481.6 l/min
0...753.3 l/min
0...1176.6 l/min
0...1990 l/min
0...3015 l/min
0...4711.6 l/min
0...7361.6 l/min
0...10601.6 l/min
0...18833.3 l/min

Technical data

 